可降解AZ31镁合金支架在兔腹主动脉的降解性能研究
介入放射学杂志2010年4月第19卷第4期JIn!!T!!咝旦型i!!垫!Q!!型:!!!堕!.璺
一315一
实验研究Experimental
research
可降解AZ31镁合金支架在兔腹主动脉的
降解性能研究
李海伟,韩洪波,
【摘要1
徐克,谭丽丽,
杨柯,刘静,
张炳春,
夏永辉,
郑
丰,
洪铎,
颜廷亭
目的初步评价可降解AZ31镁合金支架在兔主动脉的降解性能。方法将12枚可降解
AZ3l镁合金支架置人12只新西兰大白兔腹主动脉中。分别于术后1个月(n=3)、2个月(,l=3)、3个月(r/,=3)及4个月(n=3)处死动物。对支架段血管进行X线照相及病理检查。结果
12只实验兔在随访
期存活良好。X线照相及病理显示1个月时支架形态完整,扩张完全;2个月时支架部分支杆降解断裂,失去支撑作用;3个月时大部分支架支杆降解,4个月时所有支架完全降解。支架在兔主动脉中完全降解所需时间为104.5d。结论可降解AZ31镁合金支架在兔主动脉中2个月内失去支撑作用,延长支撑时间是未来的研究方向。
【关键词】可降解镁合金支架;兔腹主动脉;降解
中图分类号:R454.4文献标志码:B文章编号:1008—794X(2010)l-04-0315-03
The
degradationperformance
ofAZ31
bioabsorbable
mal;Il艘妇m枷oy
Ke。LIU
stent
implanted
inthe
abdominalaortaofrabbitsLIHai-wei,XU
Ke,MNG
Jing,ZHANGBin-chun,X1,4Yong-
Radiology,
hui,ZHENG屁昭,HANHong-bo,踢ⅣLi-li,HONGDuo,MⅣTing-ting.Departmentof
First
Affdiated
Hospital
ofChinaMedicalUnivers毋,Shenyang
110001,China
Correspondingauthor:XUKe
【Abstract】
magnesiulu
alloy
Objective
stent
Tomake
a
1
preliminaryevaluationofthedegradationofAZ3bioabsorbable
aorta
implantedintheabdominal
stentswere
ofexperimentalrabbits.
Methods
Twelve
aortas
Az31
biodegradablemagnesiumalloyNewZealandwhite
separatelydeployedintheinfrarenalabdominal
rabbits
were
oftwelve
rabbits.Every
three
experimental
sacrificedeachtime
aortaswere
at
one,two,threeand
fourmonthsaftertheprocedureofstenting.r11lestentingsegmentoftheand
harvested.radiographod
All
sentforpathologicexamination
to
obselwethedegradableperformanceofthestent.Results
animals
stentsto
survivedformthe
were
operation
inthescheduledfoHow—upperiod.Radiographieallyandpathologically。the
stentstruts
fullyexpandedwith
fractured
degradedand
perfectshapeonemonthafterintwomonths。resultinginthe
were
theprocedure,andpartofthebegan
be
lossofitssupportingfunction.Threemonthsafterthe
implantationmoststentsestimating104.5
corroded,andinfourmonthsallthesWntsbecomecornpletelydestroyed.The
stents
timeforproducingcompletedegradationofAZ3lmagnesiumalloy
AZ3l
inrabbit’saortawasaorta
days.Conclusion
bioabsorbable
magnesiumalloy
to
stents
implantedinrabbitabdominal
stents
willlosetheirradialforceintwomonths.How
next
prolong
thefunctioningtimeofthe
implanted
isthe
research
target.(JInterventRadiol,2010,19:315-317)
magnesiumalloystent;rabbit;abdominal
【Keywords】bioabsorbable
aorta;degradation
阻塞性血管疾病(包括心脑血管及外周血管疾病)严重威胁着人类健康,而且发病率呈逐年上升趋
基金项目:高等学校博士点基金INo.2伽眄015帅11)作者单位:110001
沈阳中国医科大学附属第一医院介入放
势。血管内支架植入术已经成为冠状动脉和外周血管阻塞性疾病的主要治疗手段。但是支架的长期存留导致内膜增生严重影响支架置人术的中、远期疗效…。生物可降解血管内支架(biodegradableendova-
scalarstent,BES)具有暂时存留的特点,能在人体
射科(李海伟、徐.克、刘静、夏永辉、韩洪渡、洪铎);中国科学院金属研究所(杨柯、张炳春、郑丰、谭丽丽、颜廷亭)
通信作者:徐克
内完成使命后降解消失而备受医学界关注。镁合金支架较之其他可降解支架具有支撑力强、组织相容